Title A novel non-contact radar sensor for affective and interactive analysis.
Author Lin, Hong-Dun; Lee, Yen-Shien; Shih, Hsiang-Lan; Chuang, Bor-Nian
Journal Annu Int Conf IEEE Eng Med Biol Soc Publication Year/Month 2013
PMID 24110785 PMCID -N/A-

Currently, many physiological signal sensing techniques have been applied for affective analysis in Human-Computer Interaction applications. Most known maturely developed sensing methods (EEG/ECG/EMG/Temperature/BP etc. al.) replied on contact way to obtain desired physiological information for further data analysis. However, those methods might cause some inconvenient and uncomfortable problems, and not easy to be used for affective analysis in interactive performing. To improve this issue, a novel technology based on low power radar technology (Nanosecond Pulse Near-field Sensing, NPNS) with 300 MHz radio-frequency was proposed to detect humans\' pulse signal by the non-contact way for heartbeat signal extraction. In this paper, a modified nonlinear HRV calculated algorithm was also developed and applied on analyzing affective status using extracted Peak-to-Peak Interval (PPI) information from detected pulse signal. The proposed new affective analysis method is designed to continuously collect the humans\' physiological signal, and validated in a preliminary experiment with sound, light and motion interactive performance. As a result, the mean bias between PPI (from NPNS) and RRI (from ECG) shows less than 1ms, and the correlation is over than 0.88, respectively.